MIL-STD-202G

3.2 Optional exposure procedure for the third group. The test specimens shall be located on a test surface of known area which is located 6 ±1 inches (15 ±2.5 centimeters) below a spray nozzle which discharges .16 ±.005 gpm (0.62 ±0.02 liters per minute) of solution per in2(6.5 square centimeters) of surface area at a pressure of 20 ±5 lbs/in2 (138 ±34 kilopascal). The solvent shall be held at a temperature range of 63°C to 70°C. The specimens shall be subjected to this spray for a period of 10 minutes. After completion of the spray exposure, the specimens shall be thoroughly rinsed in water and all surfaces air-blown dry and inspected in accordance with 4.1 and 4.2 to determine the extent, if any, of deterioration that has occurred. If this optional procedure is specified, brushing of the samples in the third group is not required. If a conflict arises from the use of the spray option, the brush method of 3.1d shall be used as the referee procedure. The measurements shown in parentheses are not exact equivalents and are shown for convenience only.

3.3 Immersion test for components with marking protected by a sleeve material. Components with a protective sleeve shall be divided into three groups as specified in 3, and each group shall be subjected to testing using the solution defined for that group. Each solution shall be maintained at a temperature of 25°C ±5°C, except the solution in 2.1d shall be maintained at a temperature of 63°C to 70°C. The specimens shall be completely immersed for 3 minutes +.5, -0 minutes in the specified solution contained in the vessel specified in 2.2. Immediately following immersion, each specimen shall be tested as follows: The bristle portion of the brush specified in 2.3 shall be dipped in the solution until wetted and the specimen shall be brushed with normal hand pressure (approximately 2 to 3 ounce force applied normal to the surface) for ten strokes on the sleeve directly above the area of the marking. Immediately after brushing, the procedure shall be repeated two more times for a total of three immersions followed by brushing. The brush stroke shall be directed in a forward direction across the sleeve area above the marking. After completion of the third immersion and brushing, the specimens shall be air-blown dry. The specimens shall be inspected in accordance with 4.1 and 4.2 to determine the extent, if any, of deterioration that has occurred, including the sleeve.

3.4 Open construction parts and parts not intended for PCB mounting. Parts of open construction which are susceptible to internal damage by immersion in solvents and parts not intended for mounting on printed circuit boards, shall be divided into three groups as specified in 3, and each group shall be subjected to testing using the solution defined for that group. Each solution shall be maintained at a temperature of 25°C ±5°C, except for the solution in 2.1d, which shall be maintained at a temperature range of 63°C to 70°C, and shall be contained in the vessel specified in 2.2. Each group shall be tested as follows: The bristle portion of the brush specified in 2.3 shall be immersed in the respective solution for each group until wetted. The marking area of the specimen to be tested shall then be immediately brushed with normal hand pressure (approximately 2 to 3 ounce force applied normal to the surface) for ten strokes on the portion of the specimen where marking has been applied (test area not to exceed onehalf inch square). The brush stroke shall be directed in a forward direction across the marked surface. This test shall be repeated twice for a total of three times for each specimen. The specimens shall then be inspected in accordance with 4.1 and 4.2 to determine the extent of deterioration, if any.

4. EXAMINATION AND MEASUREMENTS

4.1Marking resistance to solvents. After subjection to the test, any specified markings which are missing in whole or in part, faded, smeared, blurred, or shifted (dislodged) to the extent that they cannot be readily identified from a distance of at least 6 inches with normal room lighting without the aid of magnification or with a viewer having a magnification no greater than 3X shall constitute failure.

4.2Component protective coating, encapsulation material and sleeve material resistance. After subjection to the test, the specimen shall be examined for cracks, separations, crazing, swelling, softening, and degradation of body material, end caps and seals if present, or any other damage or degradation that has occurred due to solvent exposure and which effects the mechanical integrity or reliability shall constitute a failure. The examination shall be made with a viewer having a magnification of 10X maximum.

METHOD 215K

8 February 2002

3

MIL-STD-202G

5.SUMMARY. The following details are to be specified in the individual specification:

a.The number of specimens to be tested.

b.Optional procedure for the third group allowed (see 3.2).

TABLE I. Summary table for resistance to solvents.

Solvent 1

Solvent 2

Solvent 3

Solvent 4

1 part (by volume) of isopropyl

This solvent has

Terpene defluxer

42 parts (by volume) of

alcohol (ACS)

been deleted.

 

water

reagent grade or TT-I-735,

 

 

1 part (by volume) of

grade A or B and 3 parts (by

 

 

propylene glycol

volume) of mineral spirits per

 

 

monomethyl ether

Mil-Prf-680, type I,

 

 

1 part (by volume) of

or three parts (by volume) of a

 

 

monoethanolamine

mixture of 80% (by volume) of

 

 

 

kerosene and 20%

 

 

 

(by volume) ethylbenzene.

 

 

 

Immersion 3 +.5, -0 minutes,

 

Immersion 3 +.5, -0

Immersion 3 +.5, -0

25°C ±5°C

 

minutes, 25°C, ±5°C

minutes, 63°C to 70°C

Brush 10 strokes (wet bristle)

 

Brush 10 strokes (wet

Brush 10 strokes (wet

2 to 3 oz.

 

bristle) 2 to 3 oz.

bristle) 2 to 3 oz.

 

 

 

 

Immersion 3 +.5, -0 minutes,

 

Immersion 3 +.5, -0

Immersion 3 +.5, -0

25°C ±5°C.

 

minutes, 25°C ±5°C.

minutes, 63°C to 70°C.

Brush 10 strokes (wet bristle)

 

Brush 10 strokes (wet

Brush 10 strokes (wet

2 to 3 oz.

 

bristle) 2 to 3 oz.

bristle) 2 to 3 oz.

 

 

 

 

Immersion 3 +.5, -0 minutes,

 

Immersion 3 +.5, -0

Immersion 3 +.5, -0

25°C ±5°C.

 

minutes, 25°C ±5°C.

minutes, 63°C to 70°C.

Brush 10 strokes (wet bristle)

 

Brush 10 strokes (wet

Brush 10 strokes (wet

2 to 3 oz.

 

bristle) 2 to 3 oz.

bristle) 2 to 3 oz.

 

 

 

 

 

 

Rinse in water

Rinse in water.

Air blow dry.

 

Air blow dry

Air blow dry.

Inspect at 3X maximum for

 

Inspect at 3X maximum

Inspect at 3X maximum

marking.

 

for marking.

for marking.

Inspect at 10X maximum for

 

Inspect at 10X maximum

Inspect at 10X maximum

part damage.

 

for part damage.

for part damage.

METHOD 215K

 

 

 

8 February 2002

 

 

 

4

MIL-STD-202G

METHOD 216

RESISTANCE TO SOLDER WAVE HEAT

NOTICE

Method 216 has been canceled effective 16 April 1973. Use the test conditions in method 210 in lieu of the test conditions of method 216. Use the table below for the applicable cross reference.

When method 216

Use method 210

is specified

 

 

 

Test condition A

Test condition C

Test condition B

Test condition D

Test condition C

Test condition E

 

 

METHOD 216

16 April 1973

1 of 1

MIL-STD-202G

METHOD 217A

PARTICLE IMPACT NOISE DETECTION (PIND)

1.PURPOSE. The purpose of this test is to detect the presence of free moving particulate contaminants within sealed cavity devices. This test method is specifically directed toward relays and other devices where internal mechanism noise makes rejection exclusively by threshold level impractical. This test method also may be used prior to final sealing in the manufacturing sequence as a means of eliminating loose particles from the interior of the device.

2.APPARATUS. The basic PIND system is comprised of the following components: (Substitutions of individual units or complete system may be made providing the test requirements of paragraph 3 are met.)

a.Vibration power source consisting of electrical driver audio oscillator and low frequency audio amplifier.

b.Ultrasonic sound detection system consisting of a crystal transducer with frequency response of 36 kHz - 44 kHz and associated translator preamplifier capable of converting transducer output to 20 Hz - 5 kHz.

c.Small vibration shaker or two shakers coupled together.

d.Oscilloscope, single beam, with 100 kHz minimum bandwidth capable of external synchronization.

e.Audio speaker or headset.

f.Test fixture to adapt transducer to shaker head and isolate it from external noise sources.

g.Holding fixture designed to hold flat surface of unit under test firmly against the sensing surface of the transducer.

h.Shock test fixture (see figure 217-3).

i.Calibration unit per 3.1c.

j.Random vibration generator.

3.PROCEDURE. Test equipment shall be assembled as shown on figure 217-1. System calibration, as defined in 3.1b, shall be performed at the following intervals:

a.Each time equipment is turned on.

b.Each change of operators.

c.At initial and completion of test for each group of devices.

d.Every four hours throughout testing.

Whenever system sensitivity is found to be below specified minimum, all units tested subsequent to previous acceptable calibration shall be retested. Units rejected for particle noise shall not be reworked or retested for the purpose of acceptance. Units rejected for excessive mechanism noise may be retested for the purpose of eliminating the mechanism noise.

METHOD 217A

8 February 2002

1 of 6