Материал: part03

Внимание! Если размещение файла нарушает Ваши авторские права, то обязательно сообщите нам

DICOM PS3.3 2020a - Information Object Definitions​

Page 1601​

C.36.2.2.4 RT Treatment Position Macro​

The RT Treatment Position Macro establishes a connection between the patients geometry and the treatment delivery equipment to​ define the treatment position. When used in an RT Radiation object, this treatment position is the prescribed position. When used in​ an RT Radiation Record object, this treatment position is the record of the actual position.​

Table C.36.2.2.4-1. RT Treatment Position Macro Attributes​

Attribute Name​

Tag​

Type​

Patient Orientation Code​

(0054,0410)​

1​

Sequence​

 

 

>Include Table 8.8-1 “Code Sequence Macro Attributes”.​

 

>Patient Orientation Modifier​

(0054,0412)​

1C​

Code Sequence​

 

 

>>Include Table 8.8-1 “Code Sequence Macro Attributes”.​

Patient Equipment​

(3010,0030)​

1​

Relationship Code Sequence​

 

 

>Include Table 8.8-1 “Code Sequence Macro Attributes”.​

 

Patient Setup UID​

(300A,0650)​

1C​

Referenced RT Patient Setup​ (300A,0632)​

1C​

Sequence​

 

>Include Table 10-11 “SOP Instance Reference Macro Attributes”.​

Treatment Position Sequence​

(300A,063F)​

1​

>Treatment Position Index​

(300A,0606)​

1​

>Include Table 10.39-1 “Patient to Equipment Relationship Macro​ Attributes”.​

Attribute Description​

Sequence that describes the orientation of the patient with​ respect to gravity. See Section C.8.4.6.1.1 for further​ explanation.​

Only a single item shall be included in this Sequence.​

DCID 19 “Patient Orientation”.​

Patient Orientation Modifier.​

Required if needed to fully specify the orientation of the​ patient with respect to gravity.​

Only a single Item shall be included in this Sequence.​

DCID 20 “Patient Orientation Modifier”.​

Sequence describing the orientation of the patient with​ respect to equipment.​

Only a single Item shall be included in this Sequence.​

DCID 21 “Patient Equipment Relationship”.​

Identifies a conceptual patient setup that may or may not​ be realized by one or more RT Patient Setup Instances.​

Required if Referenced RT Patient Setup Sequence​ (300A,0632) is present. May be present otherwise.​

References the RT Patient Setup SOP Instance that was​ used as the setup instruction for the patient prior to delivery​ of the radiation.​

RequirediftherewasaPatientSetupSOPInstancedefined​ providing the instructions to the delivery system.​

Only a single Item shall be included in this Sequence.​

Patient positions during treatment, being prescribed or​ recorded.​

One or more Items shall be included in this Sequence.​

Index of this Item in this Sequence.​

The value shall start at 1 and increase monotonically by 1.​

DCID 9553 “Treatment Points”.​

C.36.2.2.5 RT Control Point General Macro​

This Macro specifies the base Attributes for the definition of an RT Radiation Control Point.​

- Standard -​

Page 1602​ DICOM PS3.3 2020a - Information Object Definitions​

Table C.36.2.2.5-1. RT Control Point General Macro Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

RT Control Point Index​ (300A,0600)​

1​

The index of the RT Control Point within the Sequence where this Macro is​

 

 

 

included.​

 

 

 

RT Control Points shall be executed in the order of the RT Control Point Index.​

 

 

 

The value shall start at 1 and increase monotonically by 1 within the Sequence​

 

 

 

where this Macro is included.​

Cumulative Meterset​

(300A,063C)​

1C​

Meterset at the RT Control Point.​

 

 

 

The units are specified by Radiation Dosimeter Unit Sequence (300A,0658).​

 

 

 

For the Item with RT Control Point Index equal 1, the Cumulative Meterset shall​

 

 

 

be equal to 0.0.​

 

 

 

RequiredifRTRadiationPhysicalandGeometricContentDetailFlag(300A,0638)​

 

 

 

equals FULL or IDENT_ONLY or RT Record Flag (300A,0639) equals YES and​

 

 

 

if the conditions in Section C.36.2.2.5.1.1 are satisfied.​

 

 

 

May be present otherwise only if the conditions in Section C.36.2.2.5.1.1 are​

 

 

 

satisfied.​

 

 

 

See Section C.36.2.2.5.1.3.​

Referenced Treatment​

(300A,060B)​

1C​

The valueofTreatmentPositionIndex(300A,0606)from theTreatmentPosition​

Position Index​

 

 

Sequence (300A,063F) within this IOD that this RT Control Point refers to.​

Required if the conditions in Section C.36.2.2.5.1.1 are satisfied.​

C.36.2.2.5.1 RT Control Point Attribute Concept​

The treatment-modality Modules use a common formalism to represent parameters that define the behaviour of a delivery device​ during delivery of radiation. These parameters are communicated as a sequence of values, organized as Control Points (see Sec-​ tion C.36.1.1.1) and represented as RT Control Points. The resolution of RT Control Points depends on the level of detail required to​ define the behaviour of the delivery device.​

A Control Point is a point on a timeline of a delivery process. RT Control Points are sequenced using an index number starting with​ 1, e.g., 1, 2, 3, 4. The RT Control Point parameters reflect the state of the delivery device at that point in time. The Control Point Cu-​ mulative Meterset reflects the dose that has been delivered from the beginning of the delivery process up to that point in time.​

For all beam deliveries there are at least two RT Control Points, corresponding to the start and end of delivery. E.g., for a simple​ Static Beam delivery with a constant field aperture, only two RT Control Points are needed to define the start and end, as there are​ no changes in-between. For a dynamic delivery, in which the MLC leaves are changing while radiation is delivered, the number of​ Control Points will be higher to provide enough detail to define the leaf movement with sufficient resolution to achieve the radiation​ fluence distribution expected for the prescribed dose.​

DICOM does not specify the behavior of the machine parameters between Control Points. The planning system needs to know the​ hardware-specific characteristics of the delivery system for which the plan is being created.​

C.36.2.2.5.1.1 Requirements for Changing Values within RT Control Point Sequence Attributes​

This Section specifies when individual attributes shall be present in a Sequence.​

The RT Control Point Sequence specifies a certain order of execution.​

At each RT Control Point the value of various Attributes may be specified as an explicit value (which in the case of a type 2C attribute​ may be a null value) and if absent remain at the same value as specified previously. There are physical and mechanical implications​ of specifying a new value as opposed to staying at the same value, for example gear lash, floating point jitter, etc.​

- Standard -​

DICOM PS3.3 2020a - Information Object Definitions​

Page 1603​

At the first Sequence Item in RT Control Point Sequences (i.e., with an RT Control Point Index (300A,0600) equal to 1) all Attributes​ affected by this Section shall be present if applicable conditions are met.​

For Sequence Items other than the first Sequence Item, Attributes shall be present if applicable conditions are met and the value is​ different from the previously populated value for the same Attribute (in the case of a type 2C attribute, a null value is considered as​ a value). The previously populated value is the value from the Item where the Attribute was present with the greatest value of RT​ Control Point Index (300A,0600) less than the value of the RT Control Point Index (300A,0600) in the current Item.​

This means that for an Item in which an Attribute is absent, the application stays at the value of the previously populated Item.​

For Sequences inside a RT Control Point Sequence Item, the Sequence shall be present if any of the nested Attributes affected by​ this Section differ from the corresponding previously populated Item.​

Formulti-valuedAttributes,suchasParallelRTBeamDelimiterPositions(300A,064A),allvaluesshallbepresentifanyvaluechanges.​

C.36.2.2.5.1.2 Control Point Attribute Example​

The following examples illustrate RT Control Points:​

1.​Static Beam delivery:​

 

 

RT Control Point Index (300A,0600)​

Cumulative Meterset (300A,063C)​

All other parameters​

1​

0​

<defined>​

2​

76​

<not present>​

At completion this beam delivers 76 Monitor Units using a fixed static set of treatment parameters defined in RT Control Point 1.​

2.​Arc delivery:​

 

 

 

RT Control Point Index​

Cumulative Meterset​

Source Roll Angle (300A,067A)​

All other parameters​

(300A,0600)​

(300A,063C)​

 

 

1​

0​

<initial angle>​

<defined>​

2​

56​

<final angle>​

<not present>​

At completion this delivers 56 Monitor Units while rotating the gantry from initial angle to final angle.​

3.​Dynamic delivery of two equally weighted segments:​

RT Control​

Cumulative​

Parallel RT Beam​

Parallel RT Beam​

RT Beam Limiting​

All other​

Point Index​

Meterset​

Delimiter Positions​

Delimiter​

Device Angle​

parameters​

(300A,0600)​

(300A,063C)​

(300A,064A)​

Positions(300A,064A)​

(300A,0679)​

 

 

 

X​

Y​

 

 

 

 

Referenced Device​

Referenced Device​

 

 

 

 

Index 1​

Index 2​

 

 

1​

0​

2\2​

2\2​

30​

<defined>​

2​

40​

<not present>​

4\4​

<not present>​

<not present>​

3​

80​

4\4​

<not present>​

<not present>​

<not present>​

At completion this delivers 80 Monitor Units while first increasing the Y opening and then increasing the X opening, while the​ beam limiting device angle stays fixed. For the RT Beam Limiting Device Opening Sequence (300A,0656) this results in having​ three Items for the first Control Point and only one for Control Points 2 (Referenced Device Index 2 only) and 3 (Referenced​ Device Index 1 only). See also Figure C.36.2.2.5.1-1.​

- Standard -​

Page 1604​

DICOM PS3.3 2020a - Information Object Definitions​

Control Point Sequence

 

 

Item 1

 

 

 

 

 

 

 

 

 

>RT Control Point Index

= 1

 

 

>Cumulative Meterset

= 0

 

 

RT BLD Settings Sequence

 

 

 

 

 

 

 

 

 

 

Item 1 (Jaw Pair, X)

 

 

 

 

 

>>Referenced Device Index

= 1

 

 

 

 

>>RT BLD Continuous Angle

= 30

 

 

 

 

>>Parallel RT Beam Delimiters Positions

= 2x2

 

 

 

Item 2 (Jaw Pair, Y)

 

 

 

 

 

>>Referenced Device Index

= 2

 

 

 

 

>>RT BLD Continuous Angle

= 30

 

 

 

 

>>Parallel RT Beam Delimiters Positions

= 2x2

 

Item 2

 

 

 

 

 

 

 

>RT Control Point Index

= 2

 

 

>Cumulative Meterset

= 40

 

 

RT BLD Settings Sequence

 

 

 

 

 

 

 

 

 

Item 1 (Jaw Pair, Y)

 

 

 

 

 

>>Referenced Device Index

= 2

 

 

 

 

>>Parallel RT Beam Delimiters Positions

= 4x4

 

Item 3

 

 

 

 

 

 

 

>RT Control Point Index

= 3

 

 

>Cumulative Meterset

= 80

 

 

RT BLD Settings Sequence

 

 

 

 

 

 

 

 

 

Item 1 (Jaw Pair, X)

 

 

 

 

 

>>Referenced Device Index

= 1

 

 

 

 

>>Parallel RT Beam Delimiters Positions

= 4x4

Figure C.36.2.2.5.1-1. Control Points Sub-Sequence Attribute Presence​

4.​Dynamic Delivery of two unequally weighted segments with a step change of 5 degrees in the positive direction of the Patient​ Support Angle:​

Note​

Patient Support Angle is represented by the Image to Equipment Mapping Matrix (0028,9520). The table contains the​ effective angle and not the complete matrix.​

RT Control Point​Cumulative Meterset​

Image to Equipment​

Source Roll Angle​

All other parameters​

Index (300A,0600)​

(300A,063C)​

Mapping Matrix​

(300A,067A)​

 

 

 

(0028,9520)​

 

 

1​

0​

0​

-90​

<defined>​

2​

30​

<not present>​

<not present>​

<not present>​

3​

<not present>​

5​

0​

<not present>​

4​

90​

<not present>​

<not present>​

<not present>​

At completion this delivers 90 Monitor Units. Between RT Control Point 2 and 3 the Patient Support Angle and Source Roll​ Continuous Angle are changed and no radiation is delivered.​

C.36.2.2.5.1.3 Cumulative Meterset​

The Meterset at a given Control Point is specified by Cumulative Meterset (300A,063C). This value is specified in units defined by​ RadiationDosimeterUnitSequence(300A,0658)intheRTDeliveryDeviceCommonModuleinSectionC.36.12.TheMetersetvalues​ are intended to correspond to the values produced by the primary or the single Meterset-measuring device of a RT Radiation Delivery​ Device.​

C.36.2.2.6 External Beam Control Point General Macro​

This Macro specifies the RT Control Point Attributes used to model external beam radiation.​

- Standard -​

DICOM PS3.3 2020a - Information Object Definitions​ Page 1605​

Table C.36.2.2.6-1. External Beam Control Point General Macro Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

Include Table C.36.2.2.5-1 “RT Control Point General Macro Attributes”.​

Delivery Rate​

(300A,063D)​

2C​

The intended nominal rate of delivery of the specified​

 

 

 

Cumulative Meterset (300A,063C).​

 

 

 

Required if the conditions in Section C.36.2.2.5.1.1 are​

 

 

 

satisfied.​

 

 

 

See Section C.36.2.2.5.1.​

Delivery Rate Unit​

(300A,063E)​

1C​

The unit of the Delivery Rate (300A,063D).​

Sequence​

 

 

Required if Delivery Rate (300A,063D) is present and has a​

 

 

 

 

 

 

value.​

 

 

 

See Section C.36.2.2.5.1.​

 

 

 

Only a single Item shall be included in this Sequence.​

>Include Table 8.8-1 “Code Sequence Macro Attributes”.​

CID is defined by invocation.​

BeamAreaLimitSequence​

(300A,0689)​

1C​

Area within which the treatment beam must be contained, for​

 

 

 

example when using MLC tracking for a moving target.​

 

 

 

Only a single Item shall be included in this Sequence.​

 

 

 

Required if beam shall be limited.​

 

 

 

See Section C.36.2.2.5.1.​

>Include Table 10.38-1 “Outline Definition Macro Attributes”.​

TheOutlineisdefinedontheBeamModifierDefinitionPlane.​

C.36.2.2.7 Radiation Generation Mode Macro​

The Radiation Generation Mode Macro contains Attributes required to generate radiation by a delivery device.​

Treatment devices can produce a multitude of different beams with properties such as energy spectrum, depth dose, surface dose​ and beam profile. A particular combination of such properties is referred to as a Radiation Generation Mode. Such Radiation Gener-​ ation Modes are created by the machine by using different primary electron/particle beams, flattening and scattering filters, etc., cre-​ atingaspecificphysicalandgeometricdistributionofradiation.InmanycasestheRadiationGenerationModecharacterizesthefluence​ just below the Monitor Chamber. Subsequently these primary beams may be modulated by beam modifiers such as Beam Limiting​ Devices, Wedges, Spreaders etc. While these beam modifiers are described in the Control Point Sequence, the primary beam is as-​ sumed to have fixed characteristics. In many cases, the Radiation Generation Mode will be constant throughout the radiation.​

Radiation Generation Modes specify the beam fluence. To convey content other than the beam fluence, such as annotating the role​ of the beam in the clinical process or the usage of that beam during a treatment session, annotate treatment constraints, use other​ AttributeslikeRTRadiationSetIntent(300A,0637)intheRTRadiationSetModuleandinformationprovidedbytheworkflowprotocols.​

Table C.36.2.2.7-1. Radiation Generation Mode Macro Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

Number of Radiation​

(300A,0685)​

1C​

NumberofRadiationGenerationModesdefinedintheRadiation​

Generation Modes​

 

 

Generation Mode Sequence (300A,067B).​

 

 

 

The Number shall be greater than zero.​

 

 

 

RequiredifRTRadiationPhysicalandGeometricContentDetail​

 

 

 

Flag (300A,0638) equals FULL. May be present otherwise.​

- Standard -​