Материал: part03

Внимание! Если размещение файла нарушает Ваши авторские права, то обязательно сообщите нам

Page 1066​

DICOM PS3.3 2020a - Information Object Definitions​

Attribute Name​

Tag​

Type​

Attribute Description​

>Table Head Tilt Angle​

(0018,9470)​

1​

Angle of the head-feet axis of the table in degrees relative to the​

 

 

 

horizontal plane. Positive values indicate that the head of the table is​

 

 

 

upwards.​

>Table Cradle Tilt Angle​

(0018,9471)​

1​

Angle of the left-right axis of the table in degrees relative to the​

 

 

 

horizontal plane. Positive values indicate that the left of the table is​

 

 

 

upwards.​

C.8.19.6.11.1 X-Ray Table Position Macro Attribute Description​

The Table Top Position Attributes of the Table Position Sequence (0018,9406) specify the geometrical position of the Table in the​ three spatial directions (i.e., Vertical, Longitudinal and Lateral) relative to the Table Top plane (see Figure C.8.19.6-3). The absolute​ reference point to which the Table positions are related is arbitrarily defined by the manufacturer.​

TheTableAngleAttributesoftheTablePositionSequence(0018,9406)specifytherotationandtiltoftheTableTopPlanewithrespect​ to a plane arbitrarily defined by the manufacturer (usually the horizontal plane).​

The Table Top Position Attributes allow to describe the incremental translation of the Table top between frames of the same Multi-​ frame image, and between frames of different images, provided that the Table Angles are not modified between these frames.​

When the table angles are modified between two frames, the Table Position Sequence (0018,9406) does not allow to characterize​ the relationship between the two table positions in an absolute reference coordinate system. For this purpose, the X-Ray Isocenter​ Reference System Macro has to be used.​

Note​

The incremental table translation may be used, in conjunction with the Positioner Position Sequence Attributes (0018,9405),​ for simple 2D-2D registration applications (object tracking, pixel shift…), assuming that the patient position is fixed on the​ table.Formorecomplexregistrationapplications,andinordertoproperlyhandlethechangesinthetableangles,itisrecom-​ mended to use the X-Ray Isocenter Reference System Macro Attributes.​

+ Lateral Position

+ Longitudinal Position

+ Vertical Position

Table Top Plane

Figure C.8.19.6-3. Table Position Vectors​

C.8.19.6.12 X-Ray Collimator Macro​

Table C.8.19.6-12 specifies the Attributes of the X-Ray Collimator Functional Group Macro.​

Table C.8.19.6-12. X-Ray Collimator Macro Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

CollimatorShapeSequence​

(0018,9407)​

1​

A Sequence that describes the collimator shape.​

 

 

 

Only a single Item shall be included in this Sequence.​

- Standard -​

 

DICOM PS3.3 2020a - Information Object Definitions​

Page 1067​

Attribute Name​

Tag​

Type​

Attribute Description​

 

>Collimator Shape​

(0018,1700)​

1​

Shape(s) of the collimator.​

 

 

 

 

Enumerated Values:​

 

 

 

 

RECTANGULAR​

 

 

 

 

CIRCULAR​

 

 

 

 

POLYGONAL​

 

 

 

 

This multi-valued Attribute shall contain at most one of each Enumerated​

 

 

 

Value.​

 

>Collimator Left Vertical​

(0018,1702)​

1C​

Required if Collimator Shape (0018,1700) is RECTANGULAR. Location​

Edge​

 

 

of the left edge of the rectangular collimator expressed as effective pixel​

 

 

 

column. See Section C.8.7.3.1.1 and Section C.8.19.6.12.1.​

 

>Collimator Right Vertical​

(0018,1704)​

1C​

Required if Collimator Shape (0018,1700) is RECTANGULAR. Location​

Edge​

 

 

of the right edge of the rectangular collimator expressed as effective pixel​

 

 

 

column. See Section C.8.7.3.1.1 and Section C.8.19.6.12.1.​

 

>Collimator Upper​

(0018,1706)​

1C​

Required if Collimator Shape (0018,1700) is RECTANGULAR. Location​

Horizontal Edge​

 

 

oftheupperedgeoftherectangularcollimatorexpressedaseffectivepixel​

 

 

 

row. See Section C.8.7.3.1.1 and Section C.8.19.6.12.1.​

 

>Collimator Lower​

(0018,1708)​

1C​

Required if Collimator Shape (0018,1700) is RECTANGULAR. Location​

Horizontal Edge​

 

 

oftheloweredgeoftherectangularcollimatorexpressedaseffectivepixel​

 

 

 

row. See Section C.8.7.3.1.1 and Section C.8.19.6.12.1.​

 

>Center of Circular​

(0018,1710)​

1C​

Required if Collimator Shape (0018,1700) is CIRCULAR. Location of the​

Collimator​

 

 

center of the circular collimator expressed as effective pixel row and​

 

 

 

column. See Section C.8.7.3.1.1 and Section C.8.19.6.12.1.​

 

>Radius of Circular​

(0018,1712)​

1C​

Required if Collimator Shape (0018,1700) is CIRCULAR. Radius of the​

Collimator​

 

 

circular collimator expressed as effective number of pixels along the row​

 

 

 

direction. See Section C.8.7.3.1.1 and Section C.8.19.6.12.1.​

>Vertices of the Polygonal​

(0018,1720)​

1C​

Required if Collimator Shape (0018,1700) is POLYGONAL.​

 

Collimator​

 

 

Multiple Values where the first set of two values are:​

 

 

 

 

 

row of the origin vertex;​ column of the origin vertex.​

Two or more pairs of values follow and are the effective pixel row and​ columncoordinatesoftheotherverticesofthepolygoncollimator.Polygon​ collimators are implicitly closed from the last vertex to the origin vertex​ and all edges shall be non-intersecting except at the vertices. See​ Section C.8.19.6.12.1.​

C.8.19.6.12.1 X-Ray Collimator Attributes​

The top left pixel of the image has a pixel row and column value of 1.​

C.8.19.6.13 X-Ray Isocenter Reference System Macro​

Table C.8.19.6-13 specifies the Attributes of the X-Ray Isocenter Reference System Functional Group Macro.​

Table C.8.19.6-13. X-Ray Isocenter Reference System Macro Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

Isocenter Reference System​

(0018,9462)​

1​ A Sequence that describes the Isocenter Reference Coordinate​

Sequence​

 

 

System (O, X, Y, Z).​

Only a single Item shall be included in this Sequence.​

- Standard -​

Page 1068​

DICOM PS3.3 2020a - Information Object Definitions​

Attribute Name​

Tag​

Type​

Attribute Description​

>Positioner Isocenter Primary​

(0018,9463)​

1​ Position of the X-Ray center beam in the isocenter reference​

Angle​

 

 

system in the X direction (deg).​

 

 

 

See Section C.8.19.6.13.1.2 for further explanation.​

>Positioner Isocenter Secondary​

(0018,9464)​

1​ Position of the X-Ray center beam in the isocenter reference​

Angle​

 

 

system in the Z direction (deg).​

 

 

 

See Section C.8.19.6.13.1.2 for further explanation.​

>Positioner Isocenter Detector​

(0018,9465)​

1​ Rotation of the X-Ray detector plane (deg).​

Rotation Angle​

 

 

See Section C.8.19.6.13.1.2 for further explanation.​

 

 

 

>Table X Position to Isocenter​

(0018,9466)​

1​ X position of the Table Reference Point with respect to the​

 

 

 

Isocenter (mm).​

 

 

 

See Section C.8.19.6.13.1.3 for further explanation.​

>Table Y Position to Isocenter​

(0018,9467)​

1​ Y position of the Table Reference Point with respect to the​

 

 

 

Isocenter (mm).​

 

 

 

See Section C.8.19.6.13.1.3 for further explanation.​

>Table Z Position to Isocenter​

(0018,9468)​

1​ Z position of the Table Reference Point with respect to the​

 

 

 

Isocenter (mm).​

 

 

 

See Section C.8.19.6.13.1.3 for further explanation.​

>Table Horizontal Rotation Angle​

(0018,9469)​

1​ Rotation of the table in the horizontal plane.​

 

 

 

See Section C.8.19.6.13.1.3 for further explanation.​

>Table Head Tilt Angle​

(0018,9470)​

1​ Angle of the head-feet axis of the table in degrees relative to the​

 

 

 

horizontal plane.​

 

 

 

See Section C.8.19.6.13.1.3 for further explanation.​

>Table Cradle Tilt Angle​

(0018,9471)​

1​ Angle of the left-right axis of the table in degrees relative to the​

 

 

 

horizontal plane.​

See Section C.8.19.6.13.1.3 for further explanation.​

C.8.19.6.13.1 Isocenter Reference System Attribute Description​

The Isocenter Reference System Attributes describe the 3D geometry of the X-Ray equipment composed by the X-Ray positioner​ and the X-Ray table.​

These Attributes define three coordinate systems in the 3D space:​

•​Isocenter coordinate system​

•​Positioner coordinate system​

•​Table coordinate system​

The Isocenter Reference System Attributes describe the relationship between the 3D coordinates of a point in the table coordinate​ system and the 3D coordinates of such point in the positioner coordinate system (both systems moving in the equipment), by using​ the Isocenter coordinate system that is fixed in the equipment.​

Note​

Annex FFF “Enhanced XA/XRF Encoding Examples (Informative)” in PS3.17 describes the transformations necessary to​ transpose between coordinate systems.​

- Standard -​

DICOM PS3.3 2020a - Information Object Definitions​

Page 1069​

C.8.19.6.13.1.1 Isocenter Coordinate System​

The Isocenter coordinate system (O,X,Y,Z) of the equipment is defined as follows:​ •​Origin O is on the System Isocenter​

•​+Y DOWNWARD (gravity)​

•​+X, +Z directions in the horizontal plane (gravity plane). Directions arbitrarily defined by the manufacturer​

+ Z

+ X

Gravity (downward)

+ Y

Gravity Plane

Figure C.8.19.6-4. Isocenter Coordinate System​

C.8.19.6.13.1.2 Positioner Coordinate System​

The Positioner Coordinate System (Op, Xp, Yp, Zp) is defined as follows:​ •​Origin Op, is the origin of the Isocenter coordinate system O​

•​Xp axis is parallel to the horizontal scan-lines of the detector (rows). Positive direction from left to right of the detector plane looking​ towards the source.​

•​Zp axis is parallel to the vertical scan-lines of the detector (columns). Positive direction from bottom to top of the detector plane​ looking towards the source.​

•​Yp is the axis from the isocenter to the X-Ray source. Positive direction from the Isocenter to the X-Ray Source. This axis is so-​ called the X-Ray center beam.​

X-ray Source

+ Zp

 

 

 

 

 

 

 

 

+ Yp

 

 

+ Xp

 

 

 

 

ISO

 

 

 

 

Rows

 

 

 

 

 

 

O

 

THLC

 

 

 

 

 

 

 

 

 

 

SID

 

 

 

 

 

 

 

Physical

 

 

 

 

 

Detector Area

 

 

 

 

 

 

 

 

 

Columns

Detector Plane

 

 

 

 

 

Figure C.8.19.6-5. Positioner Coordinate System​

Note​

The quantities SID and ISO are specified by the Attributes Distance Source to Detector (0018,1110) and Distance Source​ to Isocenter (0018,9402) respectively.​

- Standard -​

Page 1070​

DICOM PS3.3 2020a - Information Object Definitions​

The Positioner Coordinate System (Op, Xp, Yp, Zp) is characterized, with respect to the Isocenter Coordinate System (O, X, Y, Z), by​ two angles describing the X-Ray center beam, and a third angle describing the rotation of the X-Ray detector plane. These angles​ are relative to the Isocenter reference system, and independent from the patient position on the equipment.​

Positioner Isocenter Primary Angle (0018,9463) (so-called Ap1 in Figure C.8.19.6-6) is defined in the plane XY, as the angle between​ the plane YZ and the plane YpZ. The axis of rotation of this angle is the Z axis. Angle from -Y to +X is positive. The valid range of this​ angle is -180 to +180 degrees.​

Positioner Isocenter Secondary Angle (0018,9464) (so-called Ap2 in Figure C.8.19.6-6) is defined in the plane YpZ, as the angle of​ the X-Ray Center Beam (i.e., Yp) relative to the XY plane. The axis of rotation of this angle is perpendicular to the plane YpZ. Angle​ from the plane XY to +Z is positive. The valid range of this angle is -180 to +180 degrees.​

Positioner Isocenter Detector Rotation Angle (0018,9465) (so-called Ap3 in Figure C.8.19.6-6 and in Figure C.8.19.6-7) is defined in​ the detector plane, as the angle of the vertical scan-lines of the detector (i.e., Zp) relative to the intersection between the detector​ plane and the plane YpZ. The sign of this angle is positive clockwise when facing on to the detector plane (see Figure C.8.19.6-7).​ The valid range of this angle is -180 to +180 degrees.​

 

Ap3

Positioner Isocenter Primary Angle Ap1 = 0º

 

Positioner Isocenter Secondary Angle Ap2 = 0º

X-Ray Detector

 

XY plane

 

 

 

 

Ap2

 

 

Ap1

Positioner Isocenter

YpZ plane

Positioner Isocenter

 

Secondary Angle Ap2 = +90º

 

 

Primary Angle Ap = +90º

 

 

1

 

 

Isocenter

XZ plane

 

O

X-Ray Source

Figure C.8.19.6-6. Positioner Isocenter Angles​

- Standard -​