Материал: part03

Внимание! Если размещение файла нарушает Ваши авторские права, то обязательно сообщите нам

DICOM PS3.3 2020a - Information Object Definitions​

Page 981​

C.8.14.5.1.1.3 Image Flavor​

See Section C.8.16.1.3 for requirements, but not Defined Terms.​

TableC.8-108specifiestheDefinedTermsforMRSpectroscopyforValue3forImageType(0008,0008)andFrameType(0008,9007).​

Table C.8-108. MR Spectroscopy Image Type and Frame Type Value 3​

Defined Term Name​

Defined Term Description​

SPECTROSCOPY​

Spectroscopy​

C.8.14.5.1.1.4 Derived Pixel Contrast​

See Section C.8.16.1.4 for requirements, but not Defined Terms.​

Table C.8-109 specifies the Defined Terms for Value 4 for Image Type (0008,0008) and Frame Type (0008,9007).​

Table C.8-109. MR Spectroscopy Image Type and Frame Type Value 4​

Defined Term Name​

Defined Term Description​

ADDITION​

Created through point by point addition operation​

DIVISION​

Created through point by point division operation​

MAXIMUM​

Created through point by point maximum operation​

MEAN​

Created through point by point mean operation​

MINIMUM​

Created through point by point minimum operation​

MULTIPLICATION​

Created through point by point multiplication operation​

STD_DEVIATION​

Standard Deviation​

SUBTRACTION​

Created through point by point subtraction operation​

NONE​

Not calculated​

MIXED​

Used only as value in Image Type (0008,0008) if frames within the spectroscopy SOP​

 

Instance contain different values for value 4 in their Frame Type (0008,9007) Attribute.​

C.8.14.5.1.2 Volumetric Properties​

 

See Section C.8.16.2.1.2. No additional requirements or Defined Terms.​

C.8.14.5.1.3 Volume Based Calculation Technique Attribute​

See Section C.8.16.2.1.3 for requirements, but not Defined Terms.​

Table C.8-110 specifies the Defined Terms for the Volume Based Calculation Technique (0008,9207) Attribute.​

Table C.8-110. Volume Based Calculation Technique Attribute Values​

Defined Term Name​

Defined Term Description​

MAX_IP​

Maximum Intensity Projection​

MIN_IP​

Minimum Intensity Projection​

NONE​

Pixels not derived geometrically​

MIXED​

Used only as a value in Volume Based Calculation Technique (0008,9207) Attribute in the MR​

 

Spectroscopy Module if frames within the image SOP Instance contain different terms for the​

 

Volume Based Calculation Technique Attribute in MR Spectroscopy Frame Type Functional​

 

Group.​

- Standard -​

Page 982​

DICOM PS3.3 2020a - Information Object Definitions​

C.8.14.5.1.4 Complex Image Component​

The value of Complex Image Component (0008,9208) Attribute shall be used to indicate which component of the complex represent-​ ation of the signal is represented in the spectroscopy data.​

Table C.8-111 specifies the Defined Terms for Complex Image Component (0008,9208).​

Table C.8-111. Complex Image Component Attribute Values​

Defined Term Name​

Defined Term Description​

MAGNITUDE​

The magnitude component of the complex spectroscopy data.​

PHASE​

The phase component of the complex spectroscopy data.​

REAL​

The real component of the complex spectroscopy data.​

IMAGINARY​

The imaginary component of the complex spectroscopy data.​

COMPLEX​

The real and imaginary components of the complex spectroscopy data​

MIXED​

Used only as a value in Complex Image Component (0008,9208) in the MR Spectroscopy​

 

Module if frames within the image SOP Instance contain different values for the Complex Image​

 

Component Attribute in the MR Spectroscopy Frame Type Functional Group.​

C.8.14.5.1.5 Acquisition Contrast​

Table C.8-112 specifies the Defined Terms for Acquisition Contrast (0008,9209).​

 

Table C.8-112. Acquisition Contrast Values​

Defined Term Name​

Defined Term Description​

PROTON_DENSITY​

Proton Density weighted contrast​

T1​

T1 weighted contrast​

T2​

T2 weighted contrast​

UNKNOWN​

Value should be UNKNOWN if acquisition contrasts were combined resulting in an unknown​

 

contrast. Also this value should be used when the contrast is not known.​

MIXED​

Used only as a value in Acquisition Contrast (0008,9209) Attribute in the MR Spectroscopy​

 

Module if frames within the SOP Instance contain different values for the Acquisition Contrast​

 

Attribute in the MR Spectroscopy Frame Type Functional Group.​

C.8.15 Enhanced CT Image​

This section describes the specific Modules for the Enhanced CT Image IOD.​

C.8.15.1 CT Series Module​

The CT IODs use the General Series Module, specialized by the CT Series Module, to describe the DICOM Series Entity described​ in Section A.1.2.3, and to define what constitutes a Series for the context of CT device.​

Table C.8-113 specifies the Attributes that identify and describe general information about the CT Series.​

- Standard -​

DICOM PS3.3 2020a - Information Object Definitions​ Page 983​

Table C.8-113. CT Series Module Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

Modality​

(0008,0060)​

1​

Type of equipment that originally acquired the data used to​

 

 

 

create the images in this Series.​

 

 

 

Enumerated Values:​

 

 

 

CT​

 

 

 

See Section C.7.3.1.1.1 for further explanation.​

Referenced Performed​

(0008,1111)​

1C​

Uniquely identifies the Performed Procedure Step SOP​

Procedure Step Sequence​

 

 

Instance to which the Series is related.​

Only a single Item shall be included in this Sequence.​

Required if a Performed Procedure Step SOP Class was​ involved in the creation of this Series.​

>Include Table 10-11 “SOP Instance Reference Macro Attributes”​

C.8.15.2 Enhanced CT Image Module​

This section describes the Enhanced CT Image Module. Table C.8-114 specifies the Attributes of the Enhanced CT Image Module.​

Table C.8-114. Enhanced CT Image Module Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

Image Type​

(0008,0008)​

1​

Image characteristics. See Section C.8.16.1 and​

 

 

 

Section C.8.15.2.1.1.​

Multi-energy CT​

(0018,9361)​

3​

Indicates whether the image is created by means of Multi-energy​

Acquisition​

 

 

technique.​

Enumerated Values:​

YES​

NO​

Include Table C.8-131 “Common CT/MR Image Description​

Macro Attributes”​

Acquisition Number​ (0020,0012)​ 3​ A number identifying the single continuous gathering of data over​ a period of time that resulted in this image.​

Note​

This number is not required to be unique across SOP​

Instances in a Series. See also the description of the​

Referenced Raw Data Sequence (0008,9121).​

- Standard -​

Page 984​

DICOM PS3.3 2020a - Information Object Definitions​

Attribute Name​

Tag​

Type​

Attribute Description​

Acquisition DateTime​

(0008,002A)​

1C​

The date and time that the acquisition of data started.​

Note​

1.​Thesynchronizationofthistimewithanexternalclock​ is specified in the synchronization Module in​ Acquisition Time synchronized (0018,1800).​

2.​See Section C.7.6.16.2.2.1 for an overview of all​ acquisition related timing Attributes.​

 

 

 

Required if Image Type (0008,0008) Value 1 of this frame is​

 

 

 

ORIGINAL or MIXED and SOP Class UID is not​

 

 

 

"1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted), may be present​

 

 

 

otherwise.​

Acquisition Duration​

(0018,9073)​

1C​

The time in seconds needed to complete the acquisition of data.​

 

 

 

See Section C.7.6.16.2.2.1 for further explanation.​

 

 

 

Required if Image Type (0008,0008) Value 1 of this frame is​

 

 

 

ORIGINAL or MIXED and SOP Class UID is not​

 

 

 

"1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted), may be present​

 

 

 

otherwise.​

Referenced Raw Data​

(0008,9121)​

3​

The Raw data that was used to derive this Image.​

Sequence​

 

 

One or more Items are permitted in this Sequence.​

 

 

 

Note​

The Items of in this Sequence may identify raw data that​ has not been stored or encoded as a DICOM object. This​ allows recognition that images in different instances have​ been reconstructed from the same raw data.​

>Include Table C.17-3 “Hierarchical SOP Instance Reference​

Macro Attributes”​

Referenced Waveform​

(0008,113A)​

3​

References to waveforms acquired in conjunction with this image.​

Sequence​

 

 

These Waveforms may or may not be temporally synchronized with​

 

 

 

this image.​

 

 

 

One or more Items are permitted in this Sequence.​

>Include Table C.17-3 “Hierarchical SOP Instance Reference​

 

Macro Attributes”​

 

 

 

Referenced Image​

(0008,9092)​

1C​

Full set of Composite SOP Instances referring to image SOP​

Evidence Sequence​

 

 

Instances inside the frames of this Enhanced CT Image SOP​

 

 

 

Instance. See Section C.8.13.2.1.2 for further explanation.​

 

 

 

One or more Items shall be included in this Sequence.​

 

 

 

RequirediftheReferencedImageSequence(0008,1140)ispresent.​

>Include Table C.17-3 “Hierarchical SOP Instance Reference​

 

Macro Attributes”​

 

 

 

Source Image Evidence​

(0008,9154)​

1C​

Full set of Composite SOP Instances used as source image SOP​

Sequence​

 

 

Instances inside the frames of this Enhanced CT Image SOP​

Instance. See Section C.8.13.2.1.2 for further explanation.​

One or more Items shall be included in this Sequence.​

Required if the Source Image Sequence (0008,2112) is present.​

- Standard -​

 

DICOM PS3.3 2020a - Information Object Definitions​

Page 985​

Attribute Name​

Tag​

Type​

Attribute Description​

 

>Include Table C.17-3 “Hierarchical SOP Instance Reference​

 

 

Macro Attributes”​

 

 

 

 

Referenced Presentation​

(0008,9237)​

1C​

ReferencestoPresentationStateinstancesacquiredinconjunction​

State Sequence​

 

 

with this instance.​

 

 

 

 

Note​

 

 

 

 

May only be used to reference Presentation States​

 

 

 

belonging to the acquired data and not to reference​

 

 

 

PresentationStatesgeneratedsubsequentlysuchasduring​

 

 

 

interpretation.​

 

 

 

 

One or more Items shall be included in this Sequence.​

 

 

 

 

RequiredifPresentationStateisgeneratedduringacquisition,shall​

 

 

 

not be present otherwise.​

 

>Include Table C.17-3 “Hierarchical SOP Instance Reference​

 

 

Macro Attributes”​

 

 

 

 

Samples per Pixel​

(0028,0002)​

1​

Number of samples (planes) in this image. This value shall be 1.​

PhotometricInterpretation​

(0028,0004)​

1​

Specifies the intended interpretation of the pixel data.​

 

 

 

 

Enumerated Values:​

 

 

 

 

MONOCHROME2​

 

 

 

 

See Section C.7.6.3.1.2 for definition of this term.​

 

Bits Allocated​

(0028,0100)​

1​

Number of bits allocated for each pixel sample. Each sample shall​

 

 

 

have the same number of bits allocated. This value shall be 16.​

Bits Stored​

(0028,0101)​

1​

Number of bits stored for each pixel sample. Each sample shall​

 

 

 

have the same number of bits stored. This value shall be 12 or16.​

High Bit​

(0028,0102)​

1​

Most significant bit for pixel sample data. Each sample shall have​

 

 

 

the same high bit. Shall be one less than the value in Bits Stored​

 

 

 

(0028,0101).​

 

Content Qualification​

(0018,9004)​

1C​

Content Qualification Indicator​

 

Enumerated Values:​

PRODUCT​

RESEARCH​

SERVICE​

See Section C.8.13.2.1.1 for further explanation.​

Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.2.2"​ (Legacy Converted), may be present otherwise.​

Image Comments​

(0020,4000)​

3​

User-defined comments about the image​

- Standard -​