DICOM PS3.3 2020a - Information Object Definitions |
Page 1367 |
Attribute Name |
Tag |
Type |
Attribute Description |
|
Contributing Equipment |
(0018,A001) |
3 |
Sequence of Items containing descriptive Attributes of related equipment |
Sequence |
|
|
that has contributed to the acquisition, creation or modification of the |
|
|
|
Composite Instance. |
|
|
|
|
One or more Items are permitted in this Sequence. |
|
|
|
|
See Section C.12.1.1.5 for further explanation. |
|
>Purpose of Reference |
(0040,A170) |
1 |
Describesthepurposeforwhichtherelatedequipmentisbeingreferenced. |
Code Sequence |
|
|
Only a single Item shall be included in this Sequence. |
|
|
|
|
|
|
|
|
See Section C.12.1.1.5 for further explanation. |
|
>>Include Table 8.8-1 “Code Sequence Macro Attributes”DCID 7005 “Contributing Equipment Purposes of Reference”. |
>Manufacturer |
(0008,0070) |
1 |
ManufactureroftheequipmentthatcontributedtotheCompositeInstance. |
>Institution Name |
(0008,0080) |
3 |
InstitutionwheretheequipmentthatcontributedtotheCompositeInstance |
|
|
|
is located. |
|
>Institution Address |
(0008,0081) |
3 |
Address of the institution where the equipment that contributed to the |
|
|
|
Composite Instance is located. |
|
>Station Name |
(0008,1010) |
3 |
User defined name identifying the machine that contributed to the |
|
|
|
Composite Instance. |
|
>Institutional Department |
(0008,1040) |
3 |
Department in the institution where the equipment that contributed to the |
Name |
|
|
Composite Instance is located. |
|
>Institutional Department |
(0008,1041) |
3 |
A coded description of the type of Department or Service within the |
Type Code Sequence |
|
|
healthcare facility. |
|
|
|
|
Only a single Item is permitted in this Sequence. |
|
>>Include Table 8.8-1 “Code Sequence Macro Attributes”BCID 7030 “Institutional Departments, Units and Services”. |
|
>Operators' Name |
(0008,1070) |
3 |
Name(s) of the operator(s) of the contributing equipment. |
|
>Operator Identification |
(0008,1072) |
3 |
Identification of the operator(s) of the contributing equipment. |
Sequence |
|
|
One or more Items are permitted in this Sequence. |
|
|
|
|
|
|
|
|
The number and order of Items shall correspond to the number and order |
|
|
|
of values of Operators' Name (0008,1070), if present. |
|
>>Include Table 10-1 “Person Identification Macro |
|
|
Attributes Description” |
|
|
|
|
>Manufacturer's Model |
(0008,1090) |
3 |
Manufacturer's model name of the equipment that contributed to the |
Name |
|
|
Composite Instance. |
|
>Device Serial Number |
(0018,1000) |
3 |
Manufacturer's serial number of the equipment that contributed to the |
|
|
|
Composite Instance. |
|
>Software Versions |
(0018,1020) |
3 |
Manufacturer's designation of the software version of the equipment that |
|
|
|
contributed to the Composite Instance. See Section C.7.5.1.1.3. |
>Spatial Resolution |
(0018,1050) |
3 |
The inherent limiting resolution in mm of the acquisition equipment for |
|
|
|
high contrast objects for the data gathering and reconstruction technique |
|
|
|
chosen.IfvariableacrosstheimagesoftheSeries,thevalueattheimage |
|
|
|
center. |
|
>Date of Last Calibration |
(0018,1200) |
3 |
Date when the image acquisition device calibration was last changed in |
|
|
|
any way. Multiple entries may be used for additional calibrations at other |
|
|
|
times. See Section C.7.5.1.1.1 for further explanation. |
|
>Time of Last Calibration |
(0018,1201) |
3 |
Time when the image acquisition device calibration was last changed in |
|
|
|
anyway.Multipleentriesmaybeused.SeeSectionC.7.5.1.1.1forfurther |
|
|
|
explanation. |
|