DICOM PS3.3 2020a - Information Object Definitions​

Attribute Name​

Tag​

Type​

Attribute Description​

>>Include Table 8.8-1 “Code Sequence Macro Attributes”​

DCID 4242 “Cornea Measurement Method Descriptors”​

Keratometer Index​

(0022,1033)​

2​

The translation factor specific to each keratometer that derives a​

 

 

 

number for power from the measured radius of curvature of the​

 

 

 

cornea.​

Refractive Index of Cornea​

(0046,0117)​

1C​

Therefractivetranslationfactorspecificforthecorneawhenderiving​

 

 

 

a number of power from the measured radius of curvature of the​

 

 

 

posterior surface of cornea.​

 

 

 

RequiredifCorneaMeasurementTypeCodeSequence(0046,0116)​

 

 

 

contains an item with the value (DCM, 111759, "Posterior Cornea​

 

 

 

Surface Measurement").​

Refractive Index of Aqueous​

(0046,0118)​

1C​

Therefractivetranslationfactorspecificfortheaqueoushumorwhen​

Humor​

 

 

deriving a number of power from the measured radius of curvature​

 

 

 

of the posterior surface of cornea.​

RequiredifCorneaMeasurementTypeCodeSequence(0046,0116)​ contains an item with the value (DCM, 111759, "Posterior Cornea​ Surface Measurement").​

C.8.26 Visual Field Static Perimetry Measurements Modules​

C.8.26.1 Visual Field Static Perimetry Measurements Series Module​

Table C.8.26.1-1 specifies the Attributes that identify and describe general information about the Visual Field Static Perimetry Meas-​ urements Series.​

Table C.8.26.1-1. Visual Field Static Perimetry Measurements Series Module Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

Modality​

(0008,0060)​

1​

Type of equipment that originally acquired the data used to create the​

 

 

 

measurements in this Series.​

 

 

 

Enumerated Values:​

 

 

 

OPV​

 

 

 

See Section C.7.3.1.1.1 for further explanation.​

ReferencedPerformed​

(0008,1111)​

1C​

Uniquely identifies the Performed Procedure Step SOP Instance to which​

Procedure Step​

 

 

the Series is related.​

Sequence​

 

 

Only a single Item shall be included in this Sequence.​

 

 

 

 

 

 

Required if a Performed Procedure Step SOP Class was involved in the​

 

 

 

creation of this Series.​

>Include Table 10-11 “SOP Instance Reference Macro​

 

Attributes”​

 

 

 

Request Attributes​

(0040,0275)​

3​

Sequence that contains Attributes from the Imaging Service Request.​

Sequence​

 

 

One or more Items are permitted in this Sequence.​

 

 

 

- Standard -​

 

 

DICOM PS3.3 2020a - Information Object Definitions​

Page 1177​

Attribute Name​

Tag​

Type​

Attribute Description​

 

>IncludeTable10-9“RequestAttributesMacroAttributes”​In this Module Reason for Requested Procedure Code Sequence​ (0040,100A) has the Defined CID 4256 “Visual Field Procedure Modifier”.​

Note​

Reason for Requested Procedure Code Sequence (0040,100A)​ may be used to provide whether the test in the Series was taken​ for screening or diagnostic (quantitative) purposes.​

In this Module Scheduled Protocol Code Sequence (0040,0008) has the​

Defined CIDs 4250 and 4251.​

Note​

Scheduled Protocol Code Sequence (0040,0008) may be used​ to provide test pattern and test strategy.​

No Baseline CIDs are defined for other Attributes.​

IncludeTable10-16“PerformedProcedureStepSummary​In this Module Attribute Performed Protocol Code Sequence (0040,0260)​ Macro Attributes”​ is Type 1 and the Defined CID 4250 “Visual Field Static Perimetry Test​

Patterns” and CID 4251 “Visual Field Static Perimetry Test Strategies”.​

Note​

Performed Protocol Code Sequence (0040,0260) is used to​ provide test pattern and test strategy.​

In this Module Attribute Protocol Context Sequence (0040,0440) is Type​ 1 and the Defined CID 4256 “Visual Field Procedure Modifier”.​

Note​

Protocol Context Sequence (0040,0440) is used to provide​ whether the test in the Series were taken for screening or​ diagnostic purposes.​

No Baseline CIDs are defined for other Attributes.​

C.8.26.2 Visual Field Static Perimetry Test Parameters Module​

Table C.8.26.2-1 defines the Attributes used to represent the parameters used during a patient's visual field test.​

Table C.8.26.2-1. Visual Field Static Perimetry Test Parameters Module Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

Visual Field Horizontal Extent​

(0024,0010)​

1​

The maximum horizontal angular subtend (diameter or width)​

 

 

 

of the tested visual field, in degrees.​

Visual Field Vertical Extent​

(0024,0011)​

1​

The maximum vertical angular subtend (diameter or height) of​

 

 

 

the tested visual field, in degrees.​

Visual Field Shape​

(0024,0012)​

1​

The shape of the visual field tested.​

 

 

 

Defined Terms:​

RECTANGLE​

CIRCLE​

ELLIPSE​

- Standard -​

Page 1178​

DICOM PS3.3 2020a - Information Object Definitions​

Attribute Name​

Tag​

Type​

Attribute Description​

Screening Test Mode Code​

(0024,0016)​

1C​

Mode used to determine how the starting luminance values​

Sequence​

 

 

and expected thresholds are chosen.​

 

 

 

Only a single Item shall be included in this Sequence.​

 

 

 

RequiredifContentItemModifierSequence(0040,0441)within​

 

 

 

Performed Protocol Code Sequence (0040,0260) contains an​

 

 

 

Item with the value (360156006, SCT, "Screening"). May be​

 

 

 

present otherwise.​

>Include Table 8.8-1 “Code Sequence Macro Attributes”​

DCID 4252 “Visual Field Static Perimetry Screening Test​

 

 

 

Modes”​

MaximumStimulusLuminance​

(0024,0018)​

1​

Maximumluminanceofstimulus,incandelaspersquaremeter​

 

 

 

(cd/m²).​

Background Luminance​

(0024,0020)​

1​

Background luminance of the device, in candelas per square​

 

 

 

meter (cd/m²).​

 

 

 

Note​

 

 

 

This value is easily convertible to apostilb, which is​

 

 

 

used only in perimetry and is not a standardized unit.​

Stimulus Color Code​

(0024,0021)​

1​

Color of light stimulus presented to the patient.​

Sequence​

 

 

Only a single Item shall be included in this Sequence.​

 

 

 

>Include Table 8.8-1 “Code Sequence Macro Attributes”​

DCID 4255 “Visual Field Illumination Color”​

BackgroundIlluminationColor​

(0024,0024)​

1​

Color of the background illumination of the visual field device.​

Code Sequence​

 

 

Only a single Item shall be included in this Sequence.​

 

 

 

>Include Table 8.8-1 “Code Sequence Macro Attributes”​

DCID 4255 “Visual Field Illumination Color”​

Stimulus Area​

(0024,0025)​

1​

Area of light stimulus presented to the patient, in degrees​

 

 

 

squared.​

Stimulus Presentation Time​

(0024,0028)​

1​

The duration of time that a light stimulus is presented to a​

patient per each individual test point, in milliseconds.​ Note​

This time is the same for each stimulus presentation.​

C.8.26.3 Visual Field Static Perimetry Test Reliability Module​

Table C.8.26.3-1 defines the Attributes used to represent the test reliability parameters determined during a patient's visual field test.​

Table C.8.26.3-1. Visual Field Static Perimetry Test Reliability Module Attributes​

Attribute Name​

Tag​

Type​

Attribute Description​

Fixation Sequence​

(0024,0032)​

1​

The patient's gaze stability information during the visual field test.​

 

 

 

Only a single Item shall be included in this Sequence.​

>Fixation Monitoring Code​

(0024,0033)​

1​

The device strategy used to monitor the patient's fixation.​

Sequence​

 

 

One or more Items shall be included in this Sequence.​

 

 

 

>>Include Table 8.8-1 “Code Sequence Macro Attributes”​

DCID 4253 “Visual Field Static Perimetry Fixation Strategy”.​

- Standard -​

 

DICOM PS3.3 2020a - Information Object Definitions​

Page 1179​

Attribute Name​

Tag​

Type​

Attribute Description​

 

>Fixation Checked Quantity​

(0024,0035)​

1C​

The number of times that the patient's gaze fixation is checked.​

 

 

 

Required if Fixation Monitoring Code Sequence (0024,0033)​

 

 

 

contains an Item with the value (111844, DCM, "Blind Spot​

 

 

 

Monitoring") or (111845, DCM, "Macular Fixation Testing"). May​

 

 

 

be present otherwise.​

 

>Patient Not Properly Fixated​

(0024,0036)​

1C​

The number of times the patient's gaze is not properly fixated.​

Quantity​

 

 

Required if Fixation Monitoring Code Sequence (0024,0033)​

 

 

 

 

 

 

contains an Item with the value (111844, DCM, "Blind Spot​

 

 

 

Monitoring") or (111845, DCM, "Macular Fixation Testing"). May​

 

 

 

be present otherwise.​

 

>Excessive Fixation Losses​

(0024,0039)​

1​

Whetherthedevicewasabletodetermineexcessivefixationlosses.​

Data Flag​

 

 

Enumerated Values:​

 

 

 

 

 

 

 

 

YES​

 

 

 

 

NO​

 

>Excessive Fixation Losses​

(0024,0040)​

1C​

The number of fixation losses is outside of implementation-specific​

 

 

 

limits.​

 

 

 

 

Enumerated Values:​

 

 

 

 

YES​

 

 

 

 

NO​

 

 

 

 

Required if Excessive Fixation Losses Data Flag (0024,0039) is​

 

 

 

YES.​

 

Visual Field Catch Trial​

(0024,0034)​

1​

The reliability of the patient's responses to the visual field test.​

Sequence​

 

 

Only a single Item shall be included in this Sequence.​

 

 

 

>Catch Trials Data Flag​

(0024,0055)​

1​

Whether catch trials data were performed.​

 

 

 

 

Enumerated Values:​

 

 

 

 

YES​

 

 

 

 

NO​

 

>Negative Catch Trials​

(0024,0048)​

1C​

Totalnumberoftimesthepatient'svisualattentionwastestedusing​

Quantity​

 

 

stimuli brighter than previously seen luminance (negative catch​

 

 

 

trials).​

 

 

 

 

Required if Catch Trials Data Flag (0024,0055) is YES.​

>False Negatives Quantity​

(0024,0050)​

1C​

Total number of stimuli that were not seen by the patient but were​

 

 

 

previously seen at a lower luminance earlier in the visual field test​

 

 

 

(false negatives).​

 

 

 

 

Required if Catch Trials Data Flag (0024,0055) is YES.​

>False Negatives Estimate​

(0024,0045)​

1​

Whether the device was able to estimates false negatives.​

Flag​

 

 

Enumerated Values:​

 

 

 

 

 

YES​

NO​

- Standard -​

Page 1180​

DICOM PS3.3 2020a - Information Object Definitions​

Attribute Name​

Tag​

Type​

Attribute Description​

>False Negatives Estimate​

(0024,0046)​

1C​

Estimatedpercentageofallstimulithatwerenotseenbythepatient​

 

 

 

but were previously seen at a lower luminance earlier in the visual​

 

 

 

field test (false negative responses), as percent.​

 

 

 

Required if False Negatives Estimate Flag (0024,0045) is YES.​

>Excessive False Negatives​

(0024,0051)​

1​

Whether the device was able to determine excessive false​

Data Flag​

 

 

negatives.​

 

 

 

Enumerated Values:​

 

 

 

YES​

 

 

 

NO​

>Excessive False Negatives​

(0024,0052)​

1C​

The false negative estimate is outside of implementation-specific​

 

 

 

limits.​

 

 

 

Enumerated Values:​

 

 

 

YES​

 

 

 

NO​

 

 

 

Required if Excessive False Negatives Data Flag (0024,0051) is​

 

 

 

YES.​

>PositiveCatchTrialsQuantity​

(0024,0056)​

1C​

The total number of times the device behaved as if it was going to​

 

 

 

present a visual stimulus but did not actually present the stimulus​

 

 

 

(positive catch trials).​

 

 

 

Required if Catch Trials Data Flag (0024,0055) is YES.​

>False Positives Quantity​

(0024,0060)​

1C​

Thetotalnumberofpatientresponsesthatoccurredatatimewhen​

 

 

 

no visual stimulus was present (false positive responses).​

 

 

 

Required if Catch Trials Data Flag (0024,0055) is YES.​

>FalsePositivesEstimateFlag​

(0024,0053)​

1​

Whether the device was able to estimate false positives.​

 

 

 

Enumerated Values:​

 

 

 

YES​

 

 

 

NO​

>False Positives Estimate​

(0024,0054)​

1C​

Estimated percentage of all patient responses that occurred at a​

 

 

 

timewhennovisualstimuluswaspresent(falsepositiveresponses),​

 

 

 

as percent.​

 

 

 

Required if False Positives Estimate Flag (0024,0053) is YES.​

>Excessive False Positives​

(0024,0061)​

1​

Whetherthedevicewasabletodetermineexcessivefalsepositives.​

Data Flag​

 

 

Enumerated Values:​

 

 

 

 

 

 

YES​

 

 

 

NO​

>Excessive False Positives​

(0024,0062)​

1C​

The false positive estimate is outside of implementation-specific​

 

 

 

limit.​

Enumerated Values:​

YES​

NO​

Required if Excessive False Positives Data Flag (0024,0061) is​

YES.​

- Standard -​